GE Healthcare · CTLightSpeed2300000375520002300110084680001115223748718327
LightSpeed 4.X - GOC2
Symptom
Scout scan aborted with many errors
Error message
Error: 230000037 Unable to write active view data to the scan file. See Long Description for complete error. Wed May 4 17:08:51 2005 Error: 230000037 Host: ice Process: dataacq File: dataacq_save.cxx Line: 720 Function: Data Acquisition : Scan Data Save Scan Type: Scout Scan: 2801/1/2 Scan Seq Id: 552000 Unable to write active view data to the scan file. Bytes being saved: 200000h Save attempt: 3 Last view buffer? 0 Scan aborted. Error: 230011008 Unable to remove scan file from scan data disks. Error: 230011008 Unable to remove scan file from scan data disks Error: 230011008 Unable to remove scan file from scan data disks. Wed May 4 16:25:48 2005 Error: 230011008 Host: gsgect2 Process: scanmgr File: sdio_xfs_io_ops.c Line: 1164 Function: Data Acquisition : SBC Processing Scan Type: Axial Scan: 2800/200/3 Scan Seq Id: 468000 Unable to remove scan file from scan data disks. Scan: CT02/2800/200/3 Scan file: /raw_data/ex2800.CT02.1115223748.718327/ov.s200a3 OS error : 140 --- Remote I/O error Scan file may be deleted manually. Error: 230000030 Unable to write offset view data to the scan file. See LD for complete error Error: 230000030 Unable to write offset view data to the scan fil
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 4.X - GOC2 405L.8_H4.2M4 LightSpeed 4.X with GOC2 and MDAS 16 Gantry H4 Gantry Model # 2339985 and Global Operator Console 2 Model #'s 2366438 and 2366438-2(Hino) with HP8000 CPU (or newer) and Pegasus Recon ICE Box Product: LightSpeed16 405L.8_H4.2M4 CPU: GenuineIntel Intel(R) Xeon(TM) CPU 2.66GHz 4 2657 MHz i686 Processors Main memory size: 2048 Mbytes [root@presbyct2]# sg_map -i /dev/sg0 /dev/sda SEAGATE ST336753LW HPS2 /dev/sg1 /dev/sdb SEAGATE ST373453LW HPS1 /dev/sg2 /dev/sdc SONY SMO-F551-SD 1.05 /dev/sg3 /dev/scd0 MATSHITA DVD-RAM SW-9574S AZX8 /dev/sg4 /dev/scd1 HL-DT-ST DVD-ROM GDR8163B 0B15 [root@presbyct2]# df -h Filesystem Size Used Avail Use% Mounted on /dev/sda2 4.0G 1.7G 2.3G 41% / /dev/sda1 27M 12M 14M 44% /boot none 1008M 0 1007M 0% /dev/shm /dev/sda6 14G 3.5G 10G 25% /usr/g /dev/sda7 9.8G 753M 9.0G 8% /usr/g/sdc_image_pool /dev/sdb1 68G 28G 40G 41% /usr/g/sdc_image_pool2 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2225004-9 73GB scan disk (replaces 2225004-8 and 18 GB p/n 2225004-4 and 36 GB scan disk 2225004-5) See Long Description for Details. Part #: 2225004-9 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: 73GB 15000 RPM Scan Data Drive FRU: Yes Status: Orderable Comments: Per ECR 2046173, 212008898, 8/8/07. ** FE ASSIST ** Refer to Service Note SN07110612a For Jumper information (12/3/07 212001261) Part #: 2225004-8 Part Type: GP CONSUMABLES Part Weight: 1.61 LBR Repairable: No Returnable: No Hazardous Material: N Description: HARD DRIVE 73GB 15KRPM SCAN DATA FRU: Yes Status: Orderable Comments: Added per ECR 2031409, 7/14/06, ms11892. ** FE ASSIST ** If no stock of the 2225004-8 order 2225004-9 as the replacement. Refer to Service Note SN07110612a (212001261 12/3/07)
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.