AMX 4 - SN3702 Curling Fishpaper causing 439 Errors
Symptom
AMX 4 - SN3702 Curling Fishpaper Potential 439 Errors AMX-4 Curling fishpaper can cause interference with tap relays. 439 errors can occur. This problem can also cause 200 series drive errors when there is curling fishpaper by the cassette drawer.
Error message
439 Generator Fault
Applies to
- Product
- AMX4 Portables AMX4+
- Area
- X-ray R&F
Full cause & step-by-step resolution