AMX 4 - Error 214 & 216 Earth Connection at Plug J7 on the Drive Power Amplifier Intermittently Floating
Symptom
AMX 4 - Error 214 & 216 Earth Connection at Plug J7 on the Drive Power Amplifier Intermittently Floating This problem is known to have occured on at least two units. Intermittent drive stall errors which had not been cured by changing both boards. Found that the earth connection at plug J7 on the Drive Power Amplifier was intermittently floating.
Error message
Error 214 Drive Fault AMX 4 Error 216 Drive Fault AMX 4
Applies to
- Product
- AMX4 AMX4+ Portables
- Area
- X-ray R&F
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray R&F issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required