GE Healthcare · X-ray R&F

AMX 4 - Error 214 & 216 Earth Connection at Plug J7 on the Drive Power Amplifier Intermittently Floating

Symptom

AMX 4 - Error 214 & 216 Earth Connection at Plug J7 on the Drive Power Amplifier Intermittently Floating This problem is known to have occured on at least two units. Intermittent drive stall errors which had not been cured by changing both boards. Found that the earth connection at plug J7 on the Drive Power Amplifier was intermittently floating.

Error message

Error 214 Drive Fault AMX 4

Error 216 Drive Fault AMX 4

Applies to

Product
AMX4 AMX4+ Portables
Area
X-ray R&F