AMX 4 - Error 211, 214, 215 Left Drive Motor Issue
Symptom
AMX 4 - Error 211, 214, 215 Left Drive Motor Unit would not drive. Error codes 211, 214, 215. When drive was attempted the xray exposure tone and light came on. FE found the Left drive motor AMX2A1MTR1 smoked and fuse AMX1A5F1 blown. Left is referenced by looking at the drive handle facing the cassette drawer.
Error message
Error 214 Drive Fault AMX 4 Error 215 Drive Fault AMX 4 Error 211 Drive Fault AMX 4
Applies to
- Product
- AMX4 AMX4+ Portables
- Area
- X-ray R&F
Full cause & step-by-step resolution