MVP3 Phase Generator L4 KVP calibration program default values

Symptom

MVP3 Phase generator, what are the L4 KVP calibration program default values In the MVP3 Phase Generator, the factory default, or initial values for coefficients AA and BB in the L4 kVp Calibration Program, are not published in the service manuals. If the L4 program is not done correctly, it may result in wrong values for AA and BB such as: AA=000 and BBB=999. Reload the default values for AA and BB and repeat the procedure for the L4 program.

Applies to

Product
MVP 3 Phase
Area
X-ray R&F