MVP3 Error 46 at Technic
Symptom
MVP3 Generator will not terminate exposure by Forced Commutation. Error code 46 is displayed on the console. MVP Three Phase Generator will not terminate exposure by Forced Commutation. Error code 46 is displayed on the console. This problem occurs at all mA stations and at low or high kVp. The charge across forced commutation capacitor C5 rises during prep from about 160 VDC to 205 VDC very slowly (several seconds).
Error message
MVP3 Error 46
Applies to
- Product
- MVP 3 Phase
- Area
- X-ray R&F
Full cause & step-by-step resolution