GE Healthcare · CTLightSpeedVCT260133176
LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
Scan hardware reset failed - Did not detect an encoder index pulse See LD for details Scan hardware reset failed with Error: 260133172 Aperture failed to move to the requested position, Error: 260133176 Did not detect an encoder index pulse when homing CAM A, and Error: 260133128 Timed out waiting for Filter and Aperture to re-reference.
Error message
Error: 260133176 Did not detect an encoder index pulse when homing CAM A / B. See LD for more details Did not detect an encoder index pulse when homing CAM A / B. Fri Jun 9 08:41:23 2006 Error: 260133176 Host: orp Process: orp_subsys File: aperture_home.cxx Line: 721 Fri Jun 9 08:41:22 2006 Host : Collimator Ermes # : 260133176 Exception Class : Secondary Severity : Pri/Soft File : aperture_home.cxx Line# : 721 Function : No System Function Reported Scan Type : None/Unknown Scan: 310/4/0 Did not detect an encoder index pulse when homing CAM A / B
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required