MVP3 Error 17 The TIRC has not braked correctly

Symptom

MVP3 Error 17 The TIRC has not braked correctly. See long description. This error means that the TIRC has not braked. After sending the brake order, there is a wait time of 10 seconds for the TIRC to send the BRAKE code (001). When this time has elapsed, it sends the STOP code (000) unless it has sent a fresh command. This check will be carried out whenever entering the Preparation or Fluoro state when the TIRC is ordered to accelerate. This error is considered recoverable for the system and will disappear after being displayed by the console.

Error message

MVP3 Error 17: The TIRC Has Not Braked Correctly

Applies to

Product
MVP 3 Phase
Area
X-ray R&F