GE Healthcare · X-ray R&F
MVP3 Error 34 Current in Precontactor SCR During Prep State
Symptom
MVP3 Error 34 There is Current in the Precontactor SCR SCR during preparation state This error means that the Forced Commutation Precontactor SCR is on when entering Preparation or Fluoro. This check is carried out during Preparation and Fluoro. This error is considered FATAL for the system.
Error message
MVP3 ERROR 34: THERE IS CURRENT IN THE PRECONTACTOR SCR DURING PREPARATION STATE
Applies to
- Product
- MVP 3 Phase
- Area
- X-ray R&F
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray R&F issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray R&F issues
- ADx-E Advantx-E console displayed a 6C00 Error (Digital Imaging Subsystem error - Error status received in IDC Prep Reply) when the first exposure of detector bad pixel calibration was attempted; cali…
- AEC Backup Time message
- New style ion chamber does not have a preamp
- Precision 500D Table Fluoro Carriage Front Bumper is not properly adjusted, allows excessive travel toward the front of the table.
- Precision 500D Aurora, dose readout does not update when an overhead Rad Tube Exposure is made, at any technic due to abnormally low mR/mAS from the Rad Overhead XRT.
- P500D Aurora with M4 Software (May 2006 Release), Dose Readout Correct on Console, Dose Printer Prints 0.0 due to a M4 Software Bug