GE Healthcare · MRSigna22256932225704222569922256632225550
Can Not Scan. Replace the SSM.
Symptom
Can Not Scan
Error message
2225693 MDS link failure. An MDS device failed to respond to IPG. Wed Oct 29 00:14:08 1997 2225693 1 CMR1 OC0 HPC_PORT_ID PMH.m 1.29 spi_mds 503 MDS link failure. An MDS device failed to respond to IPG. If the link is broken, a separate error message will be logged. If not, a spurious error occurred. MDS timer: 0x0252 Transmitted header: 0x10 Transmitted opcode: 0x02 2225704 MDS link failure. The MDS fiber optic link appears to be broken between the RF and System cabinets. Error: 2225704 Host: lx-mr Process: NSP File: HPC_PORT_ID:spi_mds Line: 300 MDS link failure. The MDS fiber optic link appears to be broken between the RF and System cabinets. Check RF cabinet power, fiber optic connections and signal level of the fiber optic link between the cabinets. Run MDS fiber optic link diagnostics. 2225699 MDS Link Failure. An MDS device failed to respond to the IPG on a 2nd try. 2225663 ERBTEC RF Amplifier fault 84. Power Monitor Fault. 2225550 Multi-coil array driver error. A write to the multi-coil array driver coil register failed to verify See Long Description 2225550 Host: LX_MR1 Process: NSP File: HPC_PORT_ID:spi_tr_mon Line: 6300 Multi-coil array driver error.^A A write to the
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Component
- 2160200-43 SSM for LX See attached file for part information and Engineering Documentation.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.