MVP3 Error 45 See long description.
MVP3 Error 45
Symptom
MVP3 Error 45 See long description. THE POWER OFF SCR IS NOT EXCITED AT THE END OF EXPOSURE This error means that after giving the Power OFF SCR excitation order it is not working. This check is carried out for 500 uS after waiting the 100 uS from SCR excitation. In these circumstances exposure is cut off through the opening of the Safety Contactor; this action damages the Safety Contactor contacts in high power exposures. This is a FATAL error code.
Error message
MVP3 Error 45 THE POWER OFF SCR IS NOT EXCITED AT THE END OF EXPOSURE
Applies to
- Product
- MVP 3 Phase
- Area
- X-ray R&F
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray R&F issue are available to FSELIB subscribers — re-authored, web-first, searchable.