MVP3 Error 45 See long description.

MVP3 Error 45

Symptom

MVP3 Error 45 See long description. THE POWER OFF SCR IS NOT EXCITED AT THE END OF EXPOSURE This error means that after giving the Power OFF SCR excitation order it is not working. This check is carried out for 500 uS after waiting the 100 uS from SCR excitation. In these circumstances exposure is cut off through the opening of the Safety Contactor; this action damages the Safety Contactor contacts in high power exposures. This is a FATAL error code.

Error message

MVP3 Error 45 THE POWER OFF SCR IS NOT EXCITED AT THE END OF EXPOSURE

Applies to

Product
MVP 3 Phase
Area
X-ray R&F

Full cause & step-by-step resolution