Pular para o conteúdo principal
GE Healthcare · X-ray R&F

MVP3 Error 5 There is a failure of the Analog Digital Converter on the Power Unit Microcomputer Board

Symptom

MVP3 Error 5 Failure of the Analog Digital Converter on the Power Unit MIicrocomputer Board When the equipment is switched on or after a RESET. During normal operation. The test consists of measuring the voltage of the multiplexer channel 8 (U274 pin 9). This voltage should be +5 volts. It is obtained from the +15 V with a voltage divider and it reaches multiplexer U274 via an analog switch controlled by the SWAN signal. The tolerance of this voltage is +7% or -7% of 5 V. This is a FATAL error code.

Error message

MVP3 Error 5: Failure of The Analog Digital Converter On The Power Unit Microcomputer Board

Applies to

Product
MVP 3 Phase
Area
X-ray R&F

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray R&F issue are available to FSELIB subscribers — re-authored, web-first, searchable.