GE Healthcare · X-ray R&F
MVP3 Error 54 During an exposure the tube current is below the demand
Symptom
MVP3 Error 54 During an exposure the tube current is below the demand minus the tolerance. The tolerance for the mA varies depending on the mA station. For values below 100 mA, it may be more than 20% and for higher mA stations, it varies from 20% to 6%. This check is carried out during an Exposure and the check is started 12.5 mS after the Exposure start. This error is RECOVERABLE.
Error message
MVP3: Error 54
Applies to
- Product
- MVP 3 Phase
- Area
- X-ray R&F
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray R&F issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required