Symptom

MVP3 Error 14 The TIRC is not in operating condition The error may mean two things. To distinguish them, look at the CURRENT ERROR LED on the TIRC module Control board. 1) LED ON: indicates that there has been a failure in the stator current. 2) LED OFF: indicates that there has been a parity error in communication between the Power Unit Microcomputer board and the TIRC Control board. This check is done whenever the equipment is switched on. This error is considered RECOVERABLE for the system.

Error message

MVP3 Error 14

Applies to

Product
MVP 3 Phase
Component
Tirc
Area
X-ray R&F

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray R&F issue are available to FSELIB subscribers — re-authored, web-first, searchable.