LightSpeed 3.X (Ultra, Plus, QX/i) - Octane
Symptom
Scanner Hardware Reset Failed. See LD Scanner Hardware Reset Failed Hardware Failed to reset message appeared on monitor AFTER boot-up WITH Error:178568 Failed to establish socket connection to STC in 10 seconds and errors 244553 and 244554 and power up self test fails A, B, and 9 on STC CPU board The following errors were in the gesyslog Thu Jan 13 04:18:32 2005 Error: 178568 Suite: RP25 Host: SBC0 Process: fwmgr File: NBSSocket.m 1.9 Method: connectToTarget Line: 315 Failed to establish socket connection to STC in 10 seconds Thu Jan 13 04:18:32 2005 Error: 244553 Suite: RP25 Host: SBC0 Process: fwmgr File: HwSetUp.m 1.18 Method: hardReset Line: 429 The STC controller has failed to reset. Thu Jan 13 04:18:32 2005 Error: 244554 Suite: RP25 Host: SBC0 Process: fwmgr File: HwSetUp.m 1.18 Method: hardReset Line: 442 The reset failure of the OBC may be due to the reset failure of the STC. Also saw failures A, B, and 9 on the power-up test on the STC CPU .
Error message
Error: 183116 Anode mA overload detected. See Long Description for complete error Wed Jan 25 14:44:06 2005 183116 OBC Generator gen_abort.c 1.31 Function: XRAY GENERATION : Filament Selection Scan: 12087/2/1 Type: Helical Exception Level: Secondary Time: 14:44:06:200 Log Series: 202 FIL/MA fault detected: Anode mA overload detected . Desired focal spot = 0. (0 = large, 1 = small) Fault latch address = ffdf43 hex, Bit = D2 Or Desired focal spot = 1. (0 = large, 1 = small) Error: 183109 KV fault detected: Anode overcurrent See Long Description for complete error Tue Nov 30 15:38:17 2004 Error: 183109 Suite: CT01 Host: OBC Process: Generator File: gen_abort.c 1.33 Method: No Method Line: 248 Function: XRAY GENERATION : KV Loop Scan: 18763/2/1 Type: Helical Exception Level: Pri/Most Severe Time: 15:38:16:120 Log Series: 991 KV fault detected: Anode overcurrent. Fault latch address = ffcff3 hex, Bit = D4
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2154834 MA Board used on (CT/i,1.X-4.X) scanners See Long Description for more Part #: 2154834 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 1.81 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: MA CONTROL BRD./CANCEL FRU: Yes Status: Orderable Comments: FLS - MA BOARD USED ON FAST RHAPSODE FMI#25238 UPGRADES THIS BD. TO A REV.1 ONE USE LIGHTSPEED ULTRA HQT WARNING!! MATERIAL MUST BE PURCHASED FROM MFGING DO NOT GO TO VENDOR AS IT MUST BE TESTED IN MFGING BEFORE SHIPPED TO FIELD VERY VERY HIGH DOA RATE. MS08992 GVR 7/31/00GELS
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.