GE Healthcare · CTLightSpeedVCT230011073403010440290014004230019002
LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
Cannot perform any scanning. Cannot perform any scanning
Error message
Error: 230011073 STUB -SCANFILE_HEADER_callback See LD for full error Mon Jun 27 09:56:31 2005 Error: 230011073 Host: ct01 Process: headermgr File: sdio_xfs_api.c Line: 4074 STUB - SCANFILE_HEADER_callback Fri Sep 29 13:35:26 2006 Error: 230011073 Host: ct01 Process: badchanmgr File: sdio_xfs_api.c Line: 4101 STUB - SCANFILE_HEADER_callback Error detected closing information file. Shutdown processing will continue. Info file: /raw_data/info_file OS error : 2 --- No such file or directory then repeats ... Fri Sep 29 13:35:26 2006 Error: 230011073 Host: ct01 Process: scanmgr File: sdio_xfs_api.c Line: 4101 Function: Data Acquisition : SBC Processing Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 Error detected closing information file. Shutdown processing will continue. Info file: /raw_data/info_file OS error : 11 --- Resource temporarily unavailable then repeats... Fri Sep 29 13:35:46 2006 Error: 230011073 Host: ct01 Process: scanfile_xfer File: sdio_xfs_api.c Line: 4101 Function: Image Generation : SBC Processing Scan Type: Not scanning Scan: 41/3/1 Error detected closing information file. Shutdown processing will continue. Info file: /raw_data/info_file OS error : 13 --- Permi
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Component
- 5114523-2 Westville VDARC - Computer, Data Acquisition Recon Control (equivalent to 5114523) See long description for details Part #: 5114523-2 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 2.43 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: Computer, Data Acquisition Recon Control - Ed 2 FRU: Yes Status: Orderable Comments: **FE Assist** applies to CT Not Pet systems. If no stock is available order a Jarrell Darc 5159833 to replace a Westville, you MUST also order p/n 5193989 software release 06MW29.7 or LATER , and perform a LFC. See Service note SN06101010A.(2/12/07 212001261) ** Second Opinion ** Reload darc OS & APPS. For reasons unknown darc SW gets corrupt. DARC failures in system that are caused by noisy p/s fans should have a PQR written to part. IP network address should not be changed unless hospital network is on the same IP subnet. Make sure the proper DARC is ordered to match DAS/Slipring. Possible reload of DARC may fix issues. Added per ECR 2016469, this part is forward backward compatible replacement for 5114523-2. Do not use 5114523 to replace 5114523-2. ms13452 050921 Part #: 5114523 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Computer, Data Acquisition Recon Control - VDARC FRU: Yes Status: Orderable
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.