Pular para o conteúdo principal
GE Healthcare · MRSigna22255532224583222555222255512224666

Unable to Scan with Multi Coil TR Driver Faults

Symptom

Unable to Scan with Multi Coil TR Driver Faults

Error message

2225553 Multi-Coil T/R Driver Board Fault. An Open Circuit was Detected on Port 3. Thu Aug 21 12:06:10 1997 Suite: MRS1 Host: OC0 Proc: HPC_PORT_ID Error: 2225553 File: PMH.m 1.29 Method: spi_tr_mon Line: 6250 Multi-Coil T/R Driver Board Fault An open circuit was detected on port 3. Check cable to multi-coil board or check for an open pin diode.

2224583 Auto Prescan Failed, Possible Data Acquisition Failure. File: APS:aps_cf Line: 15 Auto Prescan failed, possible data acquisition failure.

2225552 Multi-Coil T/R Driver Board Fault. An Open Circuit was Detected on Port 2. Thu Aug 21 12:06:10 1997 Suite: MRS1 Host: OC0 Proc: HPC_PORT_ID Error: 2225552 File: PMH.m 1.29 Method: spi_tr_mon Line: 6240 Multi-Coil T/R Driver Board Fault An open circuit was detected on port 2. Check cable to multi-coil board or check for an open pin diode.

2225551 Multi-Coil T/R Driver Board Fault. An Open Circuit was Detected on Port 1. See Long Description Thu Aug 21 12:06:10 1997 Suite: MRS1 Host: OC0 Proc: HPC_PORT_ID Error: 2225551 File: PMH.m 1.29 Method: spi_tr_mon Line: 6230 Multi-Coil T/R Driver Board Fault An open circuit was detected on port 1. Check cable to multi-coil board or check for an op

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required