MagMon3 mounting hole locations do not match those on Granite SHeM Magnet Monitor. .
Symptom
Magnet Monitor 3 mounting hole locations do not match those on Granite SHeMs. See Resolution concept for the complete SMILe released 8/25/06. Subject: MagMon3 mounting hole locations are different than those on Granite SHeM. Effectivity: All Signa MR Systems where MagMon3 replaces Granite SHeM Magnet Monitor. Problem: MagMon3 mounting hole locations are patterned after those on Advantech SHeM Magnet Monitor. Mounting hole locations do not match those on Granite SHeM Magnet Monitor. New mounting hole locations must be drilled at sites where MagMon3 is replacing Granite SHeM.
Applies to
- Models
- Signa
- Product
- Magnet Monitor 3. Magnet Monitor 3 Magnet Monitor 3
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.