MagMon3 mounting hole locations do not match those on Granite SHeM Magnet Monitor. .
Symptom
Magnet Monitor 3 mounting hole locations do not match those on Granite SHeMs. See Resolution concept for the complete SMILe released 8/25/06. Subject: MagMon3 mounting hole locations are different than those on Granite SHeM. Effectivity: All Signa MR Systems where MagMon3 replaces Granite SHeM Magnet Monitor. Problem: MagMon3 mounting hole locations are patterned after those on Advantech SHeM Magnet Monitor. Mounting hole locations do not match those on Granite SHeM Magnet Monitor. New mounting hole locations must be drilled at sites where MagMon3 is replacing Granite SHeM.
Applies to
- Models
- Signa
- Product
- Magnet Monitor 3. Magnet Monitor 3 Magnet Monitor 3
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.