Symptom

Ring artifact after fastcals failed on cold detector. See LD for details Ring artifact appears on images after fastcals failed on cold detector.

Error message

Error: 230023046 Calibration limit check failed. Beam may be obstructed See LD for complete error Tue Mar 22 20:12:04 2005 Error: 230023046 Host: ct01 Process: dailyPrepRx File: DailyPrepUtils.c Line: 3953 Calibration limit check failed. Beam may be obstructed

Error: 230017082 Channel in non-bowtie airscan out of range - row 0, channel 10, value = 4.3 The row number can be VERY confusing. Please see Long Description for how to decode the ROW NUMBER Tue Mar 22 20:12:04 2005 Error: 230017082 Host: ct01 Process: accs File: aircal_processing.c Line: 917 Channel in non-bowtie airscan out of range - row 0, channel 10, value = 4.394680 Click Here to Decode the ROW NUMBER

Error: 230017078 Aircal failed to process scan See Long Description for complete error Tue Mar 22 20:12:04 2005 Error: 230017078 Host: ct01 Process: accs File: accs_events.cxx Line: 463 Aircal failed to process scan - Exam = 50342, Series = 1, Scan = 1, Seconds = 1111522307 Microseconds = 108580 View scanfileto check data and header information

Error: 260002451 Detector Temperature Is Low. Image quality may be affected. See LD for complete error Thu Aug 25 16:32:02 2005 Error: 260002451 Host: DAS Process: DCB Firmware 

Applies to

Models
VCT
Product
Lightspeed 1.X - Octane 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete Lightspeed 1.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config showprods|grep Apps note: if check_config does not work then the software is not up to date with all FMI's. Check FMI history and verify that all patches have been loaded since last software reload. Ref FMI 25269 List of Supported Console Types Octane with Pegasus Recon hardware. may have an 02 SBC - Run df -k to see if the sbc is partitioned. List of softwares the product can run 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3176T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.