GE Healthcare · MRSigna2241079
Mouse will not move error 2241079 Workspace Interface Module received a bad opcode
Symptom
Mouse Will Not Move Unable to Login to Boot Computer
Error message
2241079 Workspace Interface Module received a bad opcode from Workspace PC over serial link. See Long Description Error: 2241079 Host: lx-mr Process: NSP File: HPC_PORT_ID:spi_wim Line: 171 Workspace Interface Module received a bad opcode from Workspace PC over serial link. Opcode = 0x0. First, run diagnostics to test Workspace Interface Module to Workspace PC communication. If error persists, check cable connections, power cycle Workspace Interface Module and reboot Workspace PC.
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Component
- 2141978 LIGHTNING WORKSPACE INTERFACE MODULE
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.