GE Healthcare · CTLightSpeedVCT260107236260107223
44286 Hardware Stopped Scan, with Error 260107223 Gantry (Axial) could NOT find the 'C' pulse and Error
Symptom
"Scan hardware Stopped Scan" with Axial Errors. "Scan hardware Stopped Scan" with Axial Errors
Error message
Error: 260107236 Axial Motor Control (AMC) Logic will NOT Comand the DRIVE until the Gantry motion is STOPPED. See Long Description for full error Mon May 2 09:25:58 2005 Error: 260107236 Host: tgp Process: tgp_subsys File: amc_mgr.c Line: 2385 Mon May 2 09:25:58 2005 Host : Table Gantry Processor Ermes # : 260107236 Exception Class : Abort Severity : Primary File : amc_mgr.c Line# : 2385 Function : AXIAL CONTROL : Axial Control Scan Type : Helical Scan: 19/1/1 Axial Motor Control (AMC) Logic will NOT Comand the DRIVE until the Gantry motion is STOPPED. The Encoder and the Drive MUST be Sync'd/Referenced. (Cur State: 10 Cmd: 21) HW/SW Error: Retry Rx to verify probable Gantry Hardware problem. Error: 260107223 Gantry (Axial) could NOT find the 'C' pulse See LD for full error Mon May 2 09:25:53 2005 Error: 260107223 Host: tgp Process: tgp_subsys File: amc_mgr.c Line: 3551 Exception Class : Abort Severity : Primary File : amc_mgr.c Line# : 3551 Function : AXIAL CONTROL : Axial Control Scan Type : Helical Scan: 19/1/1 Gantry (Axial) could NOT find the 'C' pulse . (AMC State:4 Cmd:21 CurVel: 8.800 sec/rev). HW/SW Error: Retry Rx to verify probable Gantry Hardware problem.
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Software
- 05MW31.6 for LightSpeed VCT 7.X with Xtream FX GRE-GOC5 Console Introduced with LS VCT M3A systems produced from Oct 1st, 2005 05MW31.6 VCT 3 disk set 5149029 Includes: - Operating System Disk 5134309 - Application software Disk 5149024 - Serial Over LAN 2399297 Part Number for reference: 31.6 Class C 5149025 31.6 Class M 5149026 5149029 05MW31.6 VCT M3A Collector 3 disk set includes: 5134309 Operating System SW 5149024 Application SW 2399297 SOL Serial Over Lan SW 5149025 05MW31C.6 Class C M3A Proprietary SW 5149026 05MW31M.6 Class M M3A Proprietary SW See list of standard Options
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.