GE Healthcare · MR224018422399332239934
Getting Preamplifier Faults when using the Head Coil..
Symptom
Getting Preamplifier Faults when using the Head Coil.. Error occurs immediately on the first pulse of RF during either manual or autoprescan with the Head coil.
Error message
2240184 Phased Array bias error (short) is occurred 2239933 Coil Code has changed. 2239934 Pre amplifier error occurred, see log. Wed Jun 29 10:40:41 2005 Error: 2239934 Host: oc001 Process: NSP File: HPC_PORT_ID:spi_stop Line: 200 Pre amplifier error occurred, see log.
Applies to
- Product
- Profile 3 Ultrasparc CPU with AN8290Y gradients
- Software
- 7.7x Profile3 through 4, Contour3 through 4
- Component
- 2304198 MC Bias CBL2 May have to order FMI Kit 2305136 if 2304198 is unavailable.
- Area
- MR
Full cause & step-by-step resolution