Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT260112011

LightSpeed 7.x (VCT, VCT Select, Pro32)

Symptom

Difficulty in positioning the cradle. Cradle will not drive normally and scan aborts with errors.

Error message

Error: 260112011 Received Table Rx but the Table was NOT READY Error : 260112011 Received Table Rx but the Table was NOT READY Thu Sep 1 08:49:43 2005 Error: 260112011 Host: tgp Process: tgp_subsys File: uif_table_tilt_mgr.c Line: 3197 Thu Sep 1 08:49:44 2005 Host : Table Gantry Processor Ermes # : 260112011 Exception Class : Soft Severity : Pri/Soft File : uif_table_tilt_mgr.c Line# : 3197 Function : No System Function Reported Scan Type : Helical Scan: 8439/4/1 Received Table Rx but the Table was NOT READY for Scanning.

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Component
5135265 - VCT GT1700 Table Cradle Motor & clutch Assy See long description for FEMC info and picture 5135265 - Table Cradle Drive Motor and Clutch Assembly used on VCT GT1700 ONLY - not 2000 pics below Part #: 5135265 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: CRADLE STEPPING MOTOR and CLUTCH ASSY FRU: Yes Status: Orderable Parts Offered as: Current: 5135265 Harvested: Not Available Exchange: Not Available Comments: Not Available
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.