Pular para o conteúdo principal
GE Healthcare · CTLightSpeed2601152414889740260115217

Table does not respond to console or keypad commands

Symptom

Table does not respond to console or keypad commands

Error message

Error: 260115241 MTCB cradle invalid Pacsci drive model set. Invalid cradle current Model = xxxxxxx See LD for complete error Fri Oct 14 10:18:27 2005 Error: 260115241 Host: tgp Process: tgp_subsys File: mtcb_cmc_drv_req.c Line: 941 Fri Oct 14 10:18:28 2005 Host : Table Sub-System Ermes # : 260115241 Exception Class : Soft Severity : Pri/Soft File : mtcb_cmc_drv_req.c Line# : 941 Function : No System Function Reported Scan Type : None/Unknown Scan: 232/1/0 MTCB cradle invalid Pacsci drive model set. Invalid cradle current Model = 4889740

Error: 260115217 No response from serial device (Cradle or Elevation Drive) See LD for a sample of full error. Fri Oct 14 10:18:27 2005 Error: 260115217 Host: tgp Process: tgp_subsys File: pacsci_actions.c Line: 284 Fri Oct 14 10:18:28 2005 Host : Table Sub-System Ermes # : 260115217 Exception Class : Soft Severity : Pri/Soft File : pacsci_actions.c Line# : 284 Function : No System Function Reported Scan Type : None/Unknown Scan: 232/1/0 No response from serial device (Cradle drive) seq num 94 reply needed 0 command ff04 This same error could say "No response from serial device (Elevation Drive)"

Applies to

Models
LightSpeed
Product
LightSpeed RT4 WB O G4 G HPT WB O M4 G HPT LightSpeed RT4 with GRE and MDAS4 or GDAS4 gantry At GE Medical Systems, we’re pioneering exciting CT technology for Radiation Therapy planning. Introducing the newest addition to our LightSpeed® series – LightSpeed RT, the first multi-slice, wide-bore CT scanner with the largest field of view in the industry. With a generous 80 cm gantry opening, the LightSpeed RT bore is 10 cm larger than conventional scanners. This provides complete flexibility for patient positioning freedom – even for the most demanding oncology needs. And the 65 cm variable field of view delivers 15 additional centimeters of visual coverage, allowing clinicians to see a patient’s entire anatomy for precise radiation therapy planning. Position and precision Thin, multi-slice imaging is the extra bonus – enabling respiratory gating, providing precise target delineation and delivering higher resolution digitally reconstructed radiographs (DRRs) than ever before. All adding up to the ability for you to deliver what really matters most – better cancer care for your patients. Plus, the LightSpeed RT scanner is available with the Exact™ couch insert, enabling you to consistently replicate patient immobilization for treatment. With LightSpeed RT, the future of your Radiation Oncology department is wide open. Links to the Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2342729-3 Elevation or Cradle Drive Amplifier used on the 5.X table - same driver amp is used for both See LD for additional part information Part #: 2342729-3 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: H-POWER TABLE SERVO DRIVE, EMC EDITION 2 COMPLIANT FRU: Yes Status: Orderable Comments: **Please consider ordering the used part when it is appropriate for your customer. It will save you substantial money.** Part added per DST-HP60 16 slice RSPL. 040906 MS13452.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.