GE Healthcare · CTLightSpeed26000682150431
LightSpeed 4.X - GRE
Symptom
"Scan hardware Stopped Scan" with Axial Errors. "Scan hardware Stopped Scan" with Axial Errors
Error message
Error: 260006821 Gantry (Axial) LOOP CONTACTOR state is Still OPEN. The Loop Contactor status is NOT changing to the CLOSED state See Long Description for the full error code. Wed Jul 6 07:25:35 2005 Error: 260006821 Host: STC Process: Axial File: amc_exec.c Line: 2956 Function: AXIAL CONTROL : axial control Scan: 50431/12/1 Type: Axial Exception Level: Pri/Most Severe Time: 07:25:35:315 Log Series: 200 Gantry (Axial) LOOP CONTACTOR state is Still OPEN. The Loop Contactor status is NOT changing to the CLOSED state. (AMC State:10 Cmd:0x152b000a).
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2236254 Axial Control Board - used in H2 gantry STC 2236254 Axial Control Board 2236254 Axial Control Board Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: AXIAL CONTROL BOARD FRU: Yes Status: Orderable Comments: FLS - H2 GANTRY: COMMON TO MFG MS26561 DSM 1/7/01 COLLECTION OF DEFECTIVES START7/23/01 MS5277 FLS - PLEASE CONVERT SPH ORDER TO OVERNIGHT SHIPMENT IF PART AVAILABLE IN MEMPHIS AND IF FEDEX SERVICE (INCLUDING 8:00 AM SERVICE) MEETS FE NEED TIME CRITERIA
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.