GE Healthcare · MRSigna2244588

Collapsed Image, Bright White Line Down or Across Middle of The Image

Symptom

MRIQ Collapsed Image Bright White Line Down or Across Middle of The Image

Error message

2244588 The GP is reporting SGA Power Supply Internal Wiring Fault or Internal Power Supply Undervoltage 2244588 "The GP is reporting SGA Power Supply Internal Wiring Fault\n\ or Internal Power Supply Undervoltage. Reseat the internal\n\ SGA-PS connectors. Else, replace the SGA Power Supply." 2244588 "The GP is reporting SGA Power Supply Internal Wiring Fault or Internal Power Supply Undervoltage. Reseat the internal SGA-PS connectors. Else, replace the SGA Power Supply."

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Area
MR

Full cause & step-by-step resolution