Scanner Hardware Stopped Scan

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 260118182 (50-0201) No CAN message received within 5 sec's from See LD for error explanation Error: 260118182 (50-0201) No CAN message received within 5 sec's from Code 50 Filament Heater Errors Error code: 50-0201H Message/explanation: No CAN message received within 5 sec’s The Heater board has not received any command from the PPC kV control main software for the last 5 sec., interpreted as a loss of communication Potential cause: PPC kV control main software lost PPC kV control or Heater board CAN driver failure Bad contact on one of the pin on the CAN bus line connector Recommended action/ Troubleshooting guide: Check internal CAN bus cable between PPC kv ctl and Rotation board.

Applies to

Models
LightSpeed
Product
LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2409042 (new pn) 2325360-6 (old pn) LVPS 400 (250V) used in JEDI Aux Box of all Xi HiSpeeds and 5.X RT, 60kW 100kW LightSpeeds, Brightspeed 4/8/16 See LD for FEMC info and picture 2409042 (new pn) 2325360-6 (old pn) LVPS 400 (250V) used in JEDI Aux Box of all Xi HiSpeeds and 5.X RT, 60kW 100kW LightSpeeds picture below Note: This Item is listed as a FRU YES for the HP60 but even though the part number is the same, it is listed as a FRU NO for HP100 and RT. Part #: 2325360-6 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: LVPS 400 (250V) BOOT 2.05/APPLI 2.36 FRU: Yes Status: Orderable Comments: FLS - LVPS 400 (250V) BOOT USED ON: GENERATOR, JEDI 100 2.05/APPLI 2.36 VASC 1T ADDED PER SPCN: SUPPLIED BY GEMS INDIA 030903RF141542N1 MS13452 030908 RISK: SITE DOWN FREQUENCY: MEDIUM PER SPCN THIS PART SERVES AS REPLACEMENT FOR 2325360-5, ALL REMAINING STOCK OF -5 IS TO BE UPGRADED TO -6 VERSION. The only difference between pn 2409042 (new pn) and pn 2325360-6 (old pn) is packaging improvment . You can order pn 2325360-6 if no pole stock on pn 2409042 (6/2005 P.Liebau)
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.