Pular para o conteúdo principal
GE Healthcare · MRExciteOvation2250081

41444 Scan Stops in Middle of Scans

Symptom

Scan Stops in Middle of Scans See Long Description Hard failure during scanning, APS works ok, AP recon error watching APS window.

Error message

2250081 Z Axis Gradient Data Overrange Occurred. The GP detected an error in the incoming High Speed Data from the STIF/SRF. Wed Jun 29 06:40:35 2005 Error: 2250081 Host: lx-mr2 Process: NSP File: SCP:GP Line: 1981 Z Axis Gradient Data Overrange Occurred.^A The GP detected an error in the incoming High Speed Data from the STIF/SRF. Swap fiber optics at the Systems Cabinet. Run fiber optic diagnostics.

Applies to

Models
Excite, Ovation
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required