GE Healthcare · Nuclear MedicineMillennium

No motions on Millennium Gantry, will not Calibrate motions after resetting gantry controller, detector two NOT AT 180 d…

No motions on Millennium Gantry, will not Calibrate motions after resetting gantry controller, detector two NOT AT 180 degrees.

Symptom

No motions on Millennium Gantry, will not Calibrate motions after resetting gantry controller, detector two NOT AT 180 degrees. In some cases, typically if gantry reset and AUI has not been started yet, can move table with hhc but no rotation or radial. Checked the Gantry EUI and found VG lock in unknown status.

Applies to

Models
Millennium
Product
MG/MC/Myosight Millennium MG/MC Millennium MyoSIGHT
Component
*Gantry Gantry Gantry