LightSpeed 5.X (80/100kW)

Symptom

"Scan hardware Stopped Scan" with Axial Errors. "Scan hardware Stopped Scan" with Axial Errors

Error message

Error: 260107215 Axial Motor Control (AMC) Logic could NOT position the Gantry See long description for full error Fri Jan 6 14:13:53 2006 Error: 260107215 Host: tgp Process: tgp_subsys File: amc_mgr.c Line: 1662 Axial Motor Control (AMC) Logic could NOT position the Gantry, Retry Rx to verify probable Gantry Balance or Brake or Encoder Hardware.

Error: 260107210 Gantry (Axial) Velocity OVERSPEED - Hardware Detection. Thu Dec 9 08:37:45 2004 Host: Table Gantry Processor Ermes# : 260107210 Exception Class: Safe State Severity: Primary File: amc_isr.c Line: 527 Function: AXIAL CONTROL : Axial Control Scan Type: None/Unknown Scan 0/0/0 Gantry (Axial) Velocity OVERSPEED - Hardware Detection. Measured Velocity: 0.999 sec/rev (Reference: 1000 AMC State: 8 AMC Cmd:21) Hdw Limit: 360 msec/rev HW Error: A RESET of the Scan Hardware is required and try again. Thu Dec 9 08:37:45 2004 Error: 260107210 Host: TGP Process: tgp_subsys File: amc_isr.c Line: 527 Running diagnostic: Axial2 Functional Thu Dec 9 08:37:45 2004 Host: Table Gantry Processor Ermes# : 260107210 Exception Class: Safe State Severity: Primary File: amc_isr.c Line: 527 Function: AXIAL CONTROL : Axial Control Scan Type: None/Un

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5311936 Axial Encoder Assembly, Hub-Mounted Wheel. Used on H2, HP, VCT, and HD750. (Replaces 5182284-2 /5182284 / 2369747-2 / 2369747 / 2328307) See LD for details 5311936 Axial Encoder Assembly, Hub-Mounted Wheel. Used on H2, HP and VCT. (Replaces 5182284-2 /5182284 / 2369747-2 / 2369747 / 2328307) Part #: 5311936 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Shipping collector for Axial Encoder Assembly. FRU: Yes Status: Orderable Comments: Per ECR 2058601, 212008898, 10/8/08
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.