Pular para o conteúdo principal
GE Healthcare · MRSigna2241552

Will Not Scan with Gradient Errors

Symptom

Will Not Scan with Gradient Errors.

Error message

2241552 The GIP Reported Several PPBM Faults on the Z Axis of the GPM. Mon Apr 26 19:56:01 2004 Error: 2241552 Host: rmnf Process: NSP File: HPC_PORT_ID:GAP/GIP/GP Line: 2189 The GIP reported several PPBM Faults on the Z Axis of the GPM. Power cycle the main breaker on the GPM Power Supply to recover from this condition. If this fault occurs again, check the GPM & the GPM Power Supply.

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required