Pular para o conteúdo principal
GE Healthcare · CTLightSpeed2601152414889740

LightSpeed 5.X (60kW)

Symptom

Table does not respond to console or keypad commands

Error message

Error: 260115241 MTCB cradle invalid Pacsci drive model set. Invalid cradle current Model = xxxxxxx See LD for complete error Fri Oct 14 10:18:27 2005 Error: 260115241 Host: tgp Process: tgp_subsys File: mtcb_cmc_drv_req.c Line: 941 Fri Oct 14 10:18:28 2005 Host : Table Sub-System Ermes # : 260115241 Exception Class : Soft Severity : Pri/Soft File : mtcb_cmc_drv_req.c Line# : 941 Function : No System Function Reported Scan Type : None/Unknown Scan: 232/1/0 MTCB cradle invalid Pacsci drive model set. Invalid cradle current Model = 4889740

Applies to

Models
LightSpeed
Product
LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2342729-3 Elevation or Cradle Drive Amplifier used on the 5.X table - same driver amp is used for both See LD for additional part information Part #: 2342729-3 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: H-POWER TABLE SERVO DRIVE, EMC EDITION 2 COMPLIANT FRU: Yes Status: Orderable Comments: **Please consider ordering the used part when it is appropriate for your customer. It will save you substantial money.** Part added per DST-HP60 16 slice RSPL. 040906 MS13452.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues