PMT Tuning fails on EDI board system.

Symptom

PMT Tuning fails on EDI board system. HV DAC value is intermittently not present, issue in SW ver 4.5 b14. This issue is HV DAC value is not present in SW ver 4.5". Other name (PMT tuning problem) It was seen in manufacturing as well during first PMT tuning calibration, if the HV value is not loaded to the EDI boards The PMT tuning keeps failing and the workaround used is to power down the system, open/close AUI 4-5 attempts and then the PMT tuning start running.

Applies to

Models
Millennium
Product
MG/MC/Myosight Millennium MG/MC Millennium MyoSIGHT
Software
GenieAcq R4.5 V14 (M4) Genieacq 4.5 Build 14 (M4) 5220379 Genieacq SW R4.5 V014 M4 with accumulative patches GA450-05-1, GA450-6-1, GA450-07-0, GA450-08-1, and GA450-09-0 for Spare Part GENIE Acq R 4.5 V014 supports EDI and I/N board hardware configurations. The following computer platforms are supported by this version: SuperMicro P4SBR ( only supports the EDI detctor, reference SN43628 ) IBM x205 Dell 1400SC Dell GX200 Dell 1300. Dell 2200 and Intel AltServer are not supported. Reference CIS 5150522-100_r2