Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT2900100042046328622046331187801382426881685114147351

LightSpeed 7.x (VCT, VCT Select, Pro32)

Symptom

Cannot perform any scanning. Cannot perform any scanning

Error message

Error: 290010004 Checksum mismatch for view 1241. Expected 204632862 : Got See Long Description Tue Dec 6 11:07:48 2005 Error: 290010004 Host: darc Process: data_restore File: mdas_intf_object.cxx Line: 372 Checksum mismatch for view 1241. Expected 204632862 : Got 204633118

Buffer I/O error on device md0, logical block 78013824 seen on HP or DARC in /var/log/messages See LD for complete error Buffer I/O error on device md0, logical block 78013824 cd /var/log tail –1000 messages | more Look for any disk drive errors like these below in the HP or the DARC messages log: 16:13:48 ctcwct01 kernel: Buffer I/O error on device md0, logical block 26881685 Aug 2 12:11:33 darc kernel: Buffer I/O error on device md0, logical block 78013824

Error: 0 WARNING slow write to disk See Long Description for more details Sat Mar 4 12:03:43 2006 Error: 0 Host: darc Process: dataacq File: file_save_thread.cxx Line: 205 WARNING slow write to disk: 2.6764 sec to write 4194304bytes; scan VCT1/9992/2/1-114147351 9/400424; scan seq 6000

SCSI error : return code = 0x8000002 seen on darc in /var/log/messages seen on darc in /var/log/messages Aug 2 12:11:33 darc kernel: SCSI error : <0 0 2 0> return code = 0x

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Component
5114536-11 (replaces 5114536-10) Scan Disk Array, Hard drive with mounting sled, Seagate ST37345LC FW001 See Long Description Part #: 5114536-11 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: Scan Disk Array, Hard Drive with mounting sled, Seagate ST373455LC Firmware Revision 001 FRU: Yes Status: Orderable Comments: Per ECR 2049162, 212008898, 10/25/07. This part replaces 5114536-10. See the attachment for further details. Part #: 5114536-10 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: Scan Disk Array, Hard drive with mounting sled, Seagate ST373454LC Firmware 002 FRU: Yes Status: Orderable Comments: Added per VCT RSPL, ms13452 041216 __Set as Critical Part for Product CT_VCT, ms04278, 05Oct14 See this link to service note for details this part being obsolete and changed up to 5114536-11 : http://www.ct.med.ge.com/tech_pubs/2360027/sn/SN07102512a.pdf
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.