37292 After Fast Cal completes software begins a Z-Slope Calibration Test.

Symptom

After Fast Cal completes software begins a Z-Slope Calibration Test. See long description for details After Fast Cal completes software begins a Z-Slope Calibration Test. Screen states Generating Cal vectors, May Take 4 Minutes but the test does not complete.

Error message

Error: 230017124 Read minimum over maximum ratio from ZslopeCal.cfg is 0.550000. See LD for Full Error Error: 230017124 Read minimum over maximum ratio from ZslopeCal.cfg is 0.550000. Tue Dec 27 14:51:03 2005 Error: 230017124 Host: ct01 Process: accs File: zslope.c Line: 1910 Read minimum over maximum ratio from ZslopeCal.cfg is 0.550000.

Error: 230023009 Tube temperature before Clever Gain Air Calibration is 1046.79 degrees Celsius See Long Description for examples of tube temperature messages 230023009 Tube temperature before Clever Gain Air Calibration is 1046.79 degrees Celsius These are messages only (reporting tube temperature): Tue Nov 23 17:19:31 2004 Error: 230023009 Host: srct Process: dailyPrepRx File: StateMachineEventNotify.c Line: 447 Tube temperature before Clever Gain Air Calibration is 1046.79 degrees Celsius. another example Fri Jul 15 15:38:18 2005 Error: 230023009 Host: mvmi Process: dailyPrepRx File: StateMachineEventNotify.c Line: 418 Tube temperature before FPA Check is 716.63 degrees Celsius. another example Mon Dec 22 07:14:58 2008 230023009 mchgehdct1 dailyPrepRx StateMachineEventNotify.c 475 Tube temperature before Quality Control Scan is 114.00 degrees 

Applies to

Models
LightSpeed
Product
LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Software
05MW14.5 for all systems with GRE/Xtream consoles on HiSpeed QX/i and LightSpeed 1.X, 2.X, 3.X, 4.X, 5.X Series(QXi, PLUS, ULTRA, LS16, PRO 16, RT). Obsoleted by FMI 25365 deploy FMI 25365 It is the Service person's responsibility to use the appropriate documentation when performing and verifying service. Refer to FMI 25365: 06MW03.5 SW Update for CT-LS 1X/2X/3X/4 DIR 5161945-100, which can be found on the Distributed Technical CDROM/DVD, or online Links to any needed documentation is found in the Product concepts as an attachment called "ProductDocumentLinks"
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.