Pular para o conteúdo principal
GE Healthcare · CTLightSpeed230017801230017812000500

Scanning aborted with collimator errors

Symptom

"Scan hardware Stopped Scan" with collimator errors. "Scan hardware Stopped Scan" with collimator errors

Error message

Error: 230017801 ACCS CCal: Sweep Processing failed. Status = -1 See the Long Description for the full error code Tue Nov 30 02:33:04 2004 Error: 230017801 Host: k2_ct2_ls Process: accs File: ccal_prep.c Line: 661 ACCS CCal: Sweep Processing failed. Status = -1

Error: 230017812 ACCS CCal: Error: too much variation in ratio data. Limit 0.000500 See Long Desciption for complete error Mon May 30 05:41:49 2005 Error: 230017812 Host: cxi1 Process: accs File: collSweepProcessing.c Line: 491 ACCS CCal: Error: too much variation in ratio data. Limit 0.000500

Applies to

Models
LightSpeed
Product
Lightspeed 1.X - Octane 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete Lightspeed 1.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config showprods|grep Apps note: if check_config does not work then the software is not up to date with all FMI's. Check FMI history and verify that all patches have been loaded since last software reload. Ref FMI 25269 List of Supported Console Types Octane with Pegasus Recon hardware. may have an 02 SBC - Run df -k to see if the sbc is partitioned. List of softwares the product can run 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3176T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Software
10.5_2.8.2I_H1.3M4 for LightSpeed 1.X (QXi) with Console Assembly (2180551) (Console with O2 Computer) or (2180551-2) (Console without O2 Computer) - showprods returns LightSpeed Apps ver 10.5_2.8.2I_H1.3M4 Released via FMI 25269 The features introduced with this software include(but are not limited to) the following:Thin Twin Axial Mode (2 x .63 mm) Multi-Destination DICOM Print Capability Latest release of Iniste Interactive (IIP 2.1) Advanced Artifact Reduction (AAR) Web Browser Based Common Service Desktop - Phase II Includes a new HomePage New Log Viewer Separated Class A, C, and M software Beam-Tracking Improvement LightSpeed Apps ver 10.5_2.8.2I_H1.3M4 Released via FMI 25269 The features introduced with this software include (but are not limited to) the following: Thin Twin Axial Mode (2 x .63 mm) Multi-Destination DICOM Print Capability Latest release of Iniste Interactive (IIP 2.1) Advanced Artifact Reduction (AAR) Web Browser Based Common Service Desktop - Phase II Includes a new HomePage New Log Viewer Separated Class A, C, and M software Beam-Tracking Improvement LightSpeed QX/i Rev 1.3 System Software Collector 2260334-5 consisting of 1.3 Software CD-ROM Collector 2258825-7 and Software Rating Plate 2240213-2 1.000 PC QX/I IRIX 65 VER 6.8.A DI 2240215-2 1.000 PC QX/I IRIX 65 VER 6.8.A DI 2240216-2 1.000 PC QX/I IRIX 65U VER 6U4.5.A 2258610-6 1.000 PC LIGHTSPEED QX/I APPS SFTW
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues