Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCTHiSpeed200281001

LightSpeed 3.X (Ultra, Plus, QXi) - GRE

Symptom

Unable to view the network send queue or film composer queue

Error message

Error: 200281001 Status update: Film composer process terminated with status 0 Wed Nov 24 11:21:02 2004 Error: 200281001 Host: plsu Process: visScript File: VavFilmComposer.cxx Line: 180 Status update: Film composer process terminated with status 0 200281001 Status update: Film composer process terminated with status 0 Wed Nov 24 11:21:02 2004 Error: 200281001 Host: plsu Process: visScript File: VavFilmComposer.cxx Line: 180 Status update: Film composer process terminated with status 0

Error: 9001 Unable to set connection - server NETserver host plsu See LD for complete error Wed Nov 24 11:21:07 2004 Error: 9001 Host: plsu Process: film_composer File: application.c Line: 274 Unable to set connection - server NETserver host plsu. Tue May 17 20:23:24 2005 Error: 9001 Host: rhapsmct1 Process: netserver File: application.c Line: 215 Unable to set connection - server ARSserver host rhapsmct1.

Applies to

Models
LightSpeed, VCT, HiSpeed
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - GRE Always 4 or 8 slice MDAS. H2_P_M4 H2_P_M8 LightSpeed 3.X - GRE Method of visual identification Small rounded gantry similar in appearance to LS 2.X and 4.X. Always has an MDAS4 or MDAS8 Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da Always has an MDAS4 or MDAS8 List of Supported Console Types GOC3 Will have a DARC, SDDA and IG GOC4 Will have a DARC 2 and IG List of softwares the product can run 07MW11.10 SP3 06MW29.7 - obsolete 06MW03.5 - obsolete 05MW14.5 - obsolete 04MW44.9 - obsolete 04MW13.7 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Software
04MW44.9 for all LightSpeed systems with GRE/Xtream consoles (QXi, PLUS, ULTRA, LS16, PRO 16, RT)(1.X, 2.X, 3.X, 4.X, 5.X) Open Long Description for details. 04MW44.9 for all LightSpeed systems with GRE/Xtream consoles (QXi, PLUS, ULTRA, LS16, PRO 16, RT)(1.X, 2.X, 3.X, 4.X, 5.X) 04MW44.9A signifies Patch A has been loaded on a HP60 system. Class C Class M Application Software Set CD Part No. Ver. CD Label CD Part No. Ver. CD Label Set Part No. Apps. CD Part No Apps. CD Label showprods/ Service Desktop Intro. Date Comment 5135965 Supports Software 04MW44.9 on GRE/Xtream Class C For use on GRE/Xtream consoles on LightSpeed 5X Series H-Power 60/80/100 and MDAS-RT 5135966 Supports Software 04MW44.9 on GRE/Xtream Class M For use on GRE/Xtream consoles on LightSpeed 5X Series H-Power 60/80/100 and 5X MDAS-RT 5135967 5135962 Applications Disk 1 5135963 Applications Disk 2 5135964 DARC Applications Disk 2399296 Operating System, Ver 1.7.10 Disk 2399297 Serial Over LAN Disk 04MW44.9 May 2005 FMI 25351 For HP100, HP80, and RT Only The GRE Common Applications software supports multiple hardware configurations via a single software set. The new software version string has the following format: <software_label>.<G>_<T>_<D>_<R> The "software label" format is: YYLWxx.ZZ Where: YY = Year. For example 04. L = Location. M - Milwaukee, H - Hino, B - Beijing. Wxx = Fiscal Week. For example, W15. ZZ = Revision number for each build/reroll. Refer to Table B for an explanation of the <G>_<T>_<D>_<R> format. Table B Common Applications Sotware Naming Convention G = GANTRY TYPE T = TUBE TYPE D = DAS TYPE R = RECON HW TYPE H1 = LightSpeed 1.X P = Performix S4 = SDAS4 P = Pegasus H2 = LightSpeed 2.X, 3.X, 4.X O = OEM M4 = MDAS4 G = GRE HP = LightSpeed 5.X H = Hercules M8 = MDAS8 WB = LightSpeed 5.x RT M16 = MDAS16 QX = HighSpeed QXi G16 = GDAS16 The variables G, T, D, and R are generated dynamically based on the type of hardware configuration this software is installed on. The information for the dynamic part of the string is obtained from the INFO files which are located in /usr/g/config directory. An example string based on the above format is as follows: 04MW13.7.H2_P_M4_G The break down is as follows: This is 2004 software release at Milwaukee in Fiscal Week 13 with a revision number "7". The hardware configuration is "H2" gantry with "Performix" tube, "MDAS4", and "GRE" recon hardware. CDL Links: DIRECTION 2378261-100 GRE SOFTWARE INSTALLATION PROCEDURES Direction 2407441-100 Common Platform Q4 2004 Release Application Tips and Work-Arounds
Area
CT

Full cause & step-by-step resolution