Pular para o conteúdo principal
GE Healthcare · MRSigna2241079

CWA ID36665 System locked up with WIM and PC Errors 2241079.

Symptom

Software locked up with WIM and PC Errors

Error message

2241079 Workspace Interface Module received a bad opcode from Workspace PC over serial link. See Long Description Error: 2241079 Host: lx-mr Process: NSP File: HPC_PORT_ID:spi_wim Line: 171 Workspace Interface Module received a bad opcode from Workspace PC over serial link. Opcode = 0x0. First, run diagnostics to test Workspace Interface Module to Workspace PC communication. If error persists, check cable connections, power cycle Workspace Interface Module and reboot Workspace PC.

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues