Pular para o conteúdo principal
GE Healthcare · CTLightSpeed23000006929144370005000324400036000

LightSpeed 2.X (Plus, QX/i) - Octane

Symptom

"Scan hardware Stopped Scan" with Taxi data violation errors. "Scan hardware Stopped Scan" with Taxi data violation errors

Error message

Do Not Use - use one of the other Error: 230000069 Two, or more, unexpected events were detected during scan data transfer. The 230000069 Error has several varation see below. Sun Sep 5 11:10:31 2004 Error: 230000069 Host: darc Process: dataacq File: dip_thread.cxx Line: 396 Function: Data Acquisition : Scan Data Acquisition Scan Type: Scout Scan: 4167/1/1 Scan Seq Id: 1000 Two, or more, unexpected events were detected during scan data transfer. Bit Count Event description 3 1 TAXI modem violation 5 1 View length error Mon Dec 15 10:40:11 2008 span style="COLOR: #ff0000"> Error: 230000069 Host: darc Process: dataacq File: hddip.cxx Line: 720 Function: Data Acquisition : Scan Data Acquisition Scan Type: Helical Scan: 29144/9/3 Scan Seq Id: 37000 Two, or more, unexpected events were detected during scan data transfer. Bit Count Event description 10 1 View Data Corruption Detected: ---------------------- Bad View Details: Bit Count Event description 0 1 Checksum Field Error 1 1 View Number Field Error 2 1 Row Number Field Error 3 1 Data Length Field Error 4 1 Data Type Field Error 5 1 Magic Number Field Error 7 1 Consecutive Corrupted Segments Detected Mon Dec 22 07:16:59 2008 Error: 

Applies to

Models
LightSpeed
Product
LightSpeed 2.X (Plus, QX/i) - Octane Always S-DAS 2.4.2_H2.4M5 LightSpeed 2.X (Plus, QX/i) - Octane (Pegasus or Rhapsode IG) - Always S-DAS Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 2.4.2_H2.4M5 FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 FEATURES/OPTIONS STD/OPTION Cat # Digital Laser Camera I/F B7700PD Smart Helical (s/w only) Standard B7530TK Fast Recon (s/w only) Option B7700JF BMD (requires B7710MP or MN) Option B7710MK Smart Prep Option B7530SP Smart Score ? ? 3D Analysis (Requires D3D) Option B7540TD Navigator (s/w only) Option B7540N DentaScan (s/w only) Option B7540SD Direct 3D (Requires B7710SA) Option B7810SS HIS/RIS Option (h/w and s/w) Option B7500PN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2245896-2 HSDCD RX WITH ESD PROTECTION THIS PN REPLACES 2245896 If you have the old style ETC slip ring this part IS NOT FOR YOU. ETC uses 2167012 rf receiver See long description for more info Part #: 2245896-2 Part Type: GP CONSUMABLES Part Weight: 0.17 LBR Repairable: No Returnable: No Hazardous Material: N Description: HSDCD RX WITH ESD PROTECTION FRU: Yes Status: Orderable Comments: FLS - THIS PN REPLACES 2245896. BEFORE USING THE -2 VERSION, USE UP EXISTING STOCK OF 2245896. ADDED PER PCN 220414, BY MS20544, CSV ON 1/17/02. VC 611168 = SCHLEIFRING
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.