GE Healthcare · CTLightSpeed178574178568183113
LightSpeed 3.X (Ultra, Plus, QX/i) - GOC2
Symptom
Flash Download Fails
Error message
Error: 178574 The following controllers STC/ETC/OBC failed to report initialization status in 120 seconds See Long description for error code variations Sun Nov 27 14:18:24 2005 Error: 178568 Suite: SBCT Host: SBC0 Process: fwmgr File: NBSSocket.m 1.9 Method: connectToTarget Line: 315 Failed to establish socket connection to ETC in 10 seconds ------------------------------------------------------------- Wed Sep 8 06:20:02 2004 Error: 178568 Host: cmls Process: fwmgr File: NBSSocket.c Line: 388 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 Failed to establish socket connection to OBCR in 10 seconds. ---------------------------------------------------------------------------- Fri Dec 24 04:38:04 2004 Error: 178568 Host: ct02_sbc Process: fwmgr File: NBSSocket.c Line: 388 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 Failed to establish socket connection to STC in 10 seconds Error: 183113 KV board HVON bit high on power-up. Fri Feb 17 14:31:54 2006 Error: 183113 Host: OBC Process: Control File: Line: 235 Function: XRAY GENERATION : KV Loop Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Tim
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 3.X (Ultra, Plus, QX/i) - GOC2 This is HP CPU host running LINUX with Pegasus IG. 308L.8_H3.3M4 ~ 4 or 8 slice MDAS. ~ Could be an Ultra, Plus or QXi system. Must run swokinstall -p to verify. ~ See Long description for details . LightSpeed 3.X MDAS4 Gantry and GOC2 Gantry - 2281177-2, DAS - 2272768, Console – 2341104-2, Table - 2276776, PDU - 2269902 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2143147 KV BOARD T3202HK 2143147 KV BOARD T3202HK 2143147 KV BOARD Part #: 2143147 Part Type: GP REPAIRABLES Part Weight: 1.81 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: KV BOARD FRU: Yes Status: Orderable
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.