GE Healthcare · CTLightSpeed260006833
LightSpeed 4.X - GRE
Symptom
Gantry came to a screeching stop
Error message
Error: 260006833 Axial Motor Control (AMC) logic detected an Error with the Axial Drive LONG TERM Position Tracking. See LD for complete error Error: 260006833 Axial Motor Control (AMC) logic detected an Error with the Axial Drive LONG TERM Position Tracking. Sat Mar 25 12:52:11 2006 Error: 260006833 Host: STC Process: Axial File: amc_mon.c Line: 871 Function: AXIAL CONTROL : axial control Scan: 2592/2/1 Type: Axial Exception Level: Pri/Most Severe Time: 12:52:10:845 Log Series: 6 Axial Motor Control (AMC) logic detected an Error with the Axial Drive LONG TERM Position Tracking (TRIM) Error. Position TRIM Error: 2840. Position TRIM Tol: 1000. HW/SW Error: Retry Rx to verify probable Gantry Hardware problem.
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Software
- 05MW14.5 for all systems with GRE/Xtream consoles on HiSpeed QX/i and LightSpeed 1.X, 2.X, 3.X, 4.X, 5.X Series(QXi, PLUS, ULTRA, LS16, PRO 16, RT). Obsoleted by FMI 25365 deploy FMI 25365 It is the Service person's responsibility to use the appropriate documentation when performing and verifying service. Refer to FMI 25365: 06MW03.5 SW Update for CT-LS 1X/2X/3X/4 DIR 5161945-100, which can be found on the Distributed Technical CDROM/DVD, or online Links to any needed documentation is found in the Product concepts as an attachment called "ProductDocumentLinks"
- Area
- CT
Full cause & step-by-step resolution