Pular para o conteúdo principal
GE Healthcare · CTLightSpeed21408024294183182

LightSpeed 4.X - GRE

Symptom

Customer unable to scan with Bus voltage is present while Backup Contactor pilot relay indicates a de-energized state. Customer unable to scan with HVDC Bus voltage is present while Backup Contactor pilot relay indicates a de-energized state. Customer unable to scan with Error: 214080 HVDC Bus voltage is present while Backup Contactor pilot relay indicates a de-energized state.

Error message

Error: 214080 HVDC Bus voltage is present while Backup Contactor pilot relay indicates a de-energized state. Gentry I/O relays K251 and K245 are energized . HVDC Bus voltage : 35 Mon Mar 27 09:40:09 2006 Host : OBC Error: 214080 File: OBC diag HW Checks.c 1.36 Line: 297 Function: Diagnostics XRAY GENERATION : HVDC Backup Contactor Intlk Test Name: HV Backup Times UniqueID: 738 Iteration : 1 Exec Level: Hard Time : 24294 Error Type: Hardware HVDC Bus voltage is present while Backup Contactor pilot relay indicates a de-energized state. Gentry I/O relays K251 and K245 are energized . HVDC Bus voltage : 35 Genty I/O address FFAFBFH Bit: D4 Expected : 1 Actual : 0 May also see the following error in gesyslog: Mon Aug 16 14:39:13 2004 Error: 183182 Host: OBC Process: Generator File: gen rotor.c 1.39 Line: 498 Function: SYSTEM POWER CONTROL : HVDC Control Scan: 1935/1/1 Type: Axial Exception Level: Pri/Most Severe Time: 14:39:13:450 Log Series: 45 I/O board pilot relays are closed. PDU pilot relay not energized. BCSTAT = 0 HVDC Rail voltage = 195V Gentry I/O address: FFAFBFH Bit: D4 Expected: 1 Actual: 0 This error is flagged by the OBC Process: Generator on the HVDC Control during High V

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required