Symptom
Customer can not scan with DAS temperature errors. Customer can not scan with DAS temperature errors
Error message
Error: 260007370, DAS Temperature is LOW. See Long Description for complete error Mon Feb 20 07:43:59 2006 Error: 260007370 Host: DAS Process: DCB Firmware File: ctemptst.c Line: 490 Function: DATA ACQUISITION : Scan Data Acquisition Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 07:43:59:965 Log Series: 32 DAS Temperature is LOW. Lowest temperature was seen is on Converter Board 31. Measured: 10.5 degrees C. Acceptable Range: 15.0 to 62.0 degrees C. See Following Error Message for Converter Card List Error: 260007419 Data Acquisition System Temperature Is Low. See the LD for the complete error code Error: 260007419 Data Acquisition System Temperature Is Low. Mon Feb 6 08:37:56 2006 Error: 260007419 Host: DAS Process: DCB Firmware File: ctemptst.c Line: 474 Function: DATA ACQUISITION : Scan Data Acquisition Scan: 1277 Type: Axial Exception Level: Pri/Soft Time: 18:27:37:070 Log Series: 20 Data Acquisition System Temperature Is Low
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2336885 MDAS 4 or MDAS 8 Convertor Board, replaces 2322994-3. See Long Description for details. 2336885 MDAS 4 or MDAS 8 Convertor Card, replaces 2322994-3 (see below) Part #: 2336885 Part Type: GP CONSUMABLES Part Weight: 0.48 LBR Repairable: No Returnable: No Hazardous Material: N Description: SERVICE, MDAS CONVERTOR BD FRU: Yes Status: Orderable 4 Slice versus 8 Slice Configuration The HiSpeed QX/i system can operate 4 slice per rotation, data collections mode, whereas, the Detector and MDAS are capable of operating 4 slice or 8 slice per rotation. From a hardware perspective the main difference is in the MDAS. The 4 slice configuration uses a depopulated MDAS. This means that all EVEN slots are empty. Only ODD numbered slots contain converter boards. Data collection is still performed in the same manner using only the odd converter card slot assignments.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.