Symptom

F3 fuse on the GTCB blows whenever cradle motion is commanded

Error message

Error: 260134616 Cradle clutch lines are disconnected Cannot control cradle clutch See long description for full error Mon Jun 27 13:22:22 2005 Error: 260134616 Host: tgp Process: tgp_subsys File: gtcb_cmc_drv_req.cxx Line: 757 Mon Jun 27 13:22:22 2005 Host : Table Sub-System Ermes # : 260134616 Exception Class : Secondary Severity : Pri/Soft File : gtcb_cmc_drv_req.cxx Line# : 757 Function : No System Function Reported Scan Type : Scout Scan: 71/1/1 Cradle clutch lines are disconnected. Cannot control cradle clutch

Error: 260134619 Cradle linear scale error not cleared. Please check cable connection See Long Description for error example Fri Mar 24 13:32:41 2006 Error: 260134619 Host: tgp Process: tgp_subsys File: gtcb_cmc_drv_req.cxx Line: 885 Fri Mar 24 13:32:41 2006 Host : Table Sub-System Ermes # : 260134619 Exception Class : Secondary Severity : Pri/Soft File : gtcb_cmc_drv_req.cxx Line# : 885 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 Cradle linear scale error not cleared. Please check cable connection <This could be your problem

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Component
5114122 Cable from GTCB J11 TO J200 Positioning GT See LD for details P/N 5114122 Cable from GTCB J11 TO J200 Positioning GT -no picture or drawing available Part #: 5114122 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: GTCBJ11 TO J200 Positioning GT FRU: Yes Status: Orderable Parts Offered as: Current: 5114122 Harvested: Not Available Exchange: Not Available
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.