GE Healthcare · CTHiSpeedLightSpeed26000745650201
Fine streaks appear on images with beam tracking errors
Symptom
Fine streaks appear on images with beam tracking errors See Long Description for artifact Fine streaks appear on images with beam tracking errors
Error message
Error: 260007456 Beam Tracking Control Loop Error Exceeded The WARNING Tolerance. See LD for more details Error: 260007456 Beam Tracking Control Loop Error Exceeded The WARNING Tolerance Mon Mar 20 16:36:22 2006 Error: 260007456 Host: DAS Process: DCB Firmware File: zaxis.c Line: 578 Function: DATA ACQUISITION : Scan Data Acquisition Scan: 50201/5/19 Type: Axial Exception Level: Pri/Soft Time: 16:36:22:380 Log Series: 83 Beam Tracking Control Loop Error Exceeded The WARNING Tolerance. Tolerance : 200 umd Control Error : 319 umd Beam Side : A Side
Applies to
- Models
- HiSpeed, LightSpeed
- Product
- HiSpeed QX/i - GOC2 103L.5_HSQXIM4 103L.4_HSQXIM4 102L.1_HSQXIM3 HiSpeed QX/i - GOC2 Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types GOC2 List of softwares the product can run 103L.5_HSQXIM4 103L.4_HSQXIM4 - obsolete 102L.1_HSQXIM3 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2159578-3 collimator control board used on 1X/2X/3X/4X/5X See Long description for details Part #: 2159578-3 Part Type: GP CONSUMABLES Part Weight: 2.18 LBR Repairable: No Returnable: No Hazardous Material: N Description: COLLIMATOR CONTROL BOARD FRU: Yes Status: Orderable Comments: Not Available
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.