GE Healthcare · CTLightSpeedVCT1841891002112469
LightSpeed 1.X with S-DAS4 Gantry and GRE
Symptom
Intermittent Scan Aborts when table is moving between 1st and 2nd scouts, between helical series, or between axial series. Clicking Pause / Resume restarts scan. See long description for details. Intermittent Scan Aborts when table is moving between 1st and 2nd scouts, between helical series, or between axial series. The customer can click "Pause" and then "Resume" and continue scanning, after clicking "Resume or "Pause" it can take from 10 to 60 seconds before system will start scanning again.
Error message
Error: 184189 SW ERROR: Software Problem. Error occured when writing CF message 10021. See LD for complete error Mon Jul 11 09:51:24 2005 Error: 184189 Host: STC Process: Scan Control File: scRxMgr.c 1.48 Line: 251 Function: No System Function Reported < Scan: 12469/2/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 09:51:23:970 Log Series: 958 CF error 2103 detected on sending a CF message. SW ERROR: Software Problem. Error occured when writing CF message 10021
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 1.X - GRE H1_P_S4 Lightspeed 1.X - GRE Method of visual identification Large square gantry similar in appearance to a CTi gantry. Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console Types GOC3 Will have a DARC, SDDA and IG GOC4 Will have a DARC 2 and IG List of softwares the product can run 07MW11.10 SP3 06MW29.7 - obsolete 06MW03.5 - obsolete 05MW14.5 - obsolete 04MW44.9 - obsolete 04MW13.7 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3176T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2224296-3 Artysen board for 1X/2X/3X/4X used in the ETC/STC/OBC See Long Description for FEMC information and board locations Part #: 2224296-3 Part Type: GP REPAIRABLES Part Weight: 1.09 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: VE-GE, 8MB, Artesyn CPU Bd FRU: Yes Status: Orderable Parts Offered as: Current : 2224296-3 Harvested : Not Available Exchange : Not Available Comments: ** FE ASSIST ** Repair operations has found that 68% of the time the Artesyn board (2224296-3) is returned for repair, it has No Defect Found. Check switch settings (212001261 2/12/09)
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.