GE Healthcare · MRSigna2000023682223822222381722320962000023522232100
System fails to boot Applications software and TPS reset fails.
Symptom
System fails to boot Applications software and TPS reset fails. See long description. The system is a 1.5T LX, running 9.1 M4 software. The system will not boot applications. Looks like the Bit3 is down. Errors in the log look like TPS reset failure. Ping to tps fails as seen below: lx-mr 6# ping lx-mr-tps PING lx-mr-tps (216.33.27.2): 56 data bytes ping: sendto: Connection refused
Error message
200002368 A set of processes are deadlocked waiting for each other to start. Mon Apr 12 11:43:42 2004 Error: 200002368< Host: BaptistGEMR Process: hostMonitorbsp File: ssProcStart.c Line: 1099 A set of processes are deadlocked waiting for each other to start. svat, scn, rtcProxy, resetServer, lais, hgp, displayServer, autoview, app, WLServer, TIR, SITT, RtpHost, RTPSIServer, RIR, PPSproxy are the likely culprits. This deadlock has caused the system to shutdown! 2223822 Reset TPS was not successful. Check System Cabinet power and connections via BIT3 interface. Sat Nov 6 07:30:31 2004 Error: 2223822 Host: lx-mr1c Process: NSP File: TardisReset.cc Line: 522 Reset TPS was not successful. Check System Cabinet power and connections via BIT3 interface. 2223817 TPS is Not Responding. Please Reset TPS Wed Apr 28 16:58:01 2004 Error: 2223817 Host: GEMS Process: NSP File: TardisReset.cc Line: 398 TPS is not responding. Please reset TPS. 2232096 Internal Software Error. Failure in creating hostSpectroMgr object. Cannot continue. Thu Apr 21 07:23:59 2005 Error: 2232096 Host: lx-mr Process: spectro File: spectroMain.cc Line: 187 Internal Software Error. Failure in creating hostSpectroMgr obj
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Component
- 2118408-3 MAIN POWER SUPPLY - BACKPLANE This is the TPS/ISE Backplane Power Supply. It is referred to in Renewal Parts as MAIN POWER SUPPLY – BACKPLANE
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.