Pular para o conteúdo principal
GE Healthcare · MRSigna2241515

Will not Scan with Minigram Overvoltage faults Due to Minigram Control Board

Symptom

Will Not Scan with Gradient Errors.

Error message

2241515 The GIP reported a X Axis miniGRAM OverVoltage Fault 2241515 File: HPC_PORT_ID:GAP/GIP/GP Line: 2092 The GIP reported a X Axis miniGRAM OverVoltage Fault. Try swapping the miniGRAM Control Board, or the entire miniGRAM axis, or the miniGRAM Power Supply to isolate the problem.->

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Component
2172312 Minigram Control Board
Area
MR

Full cause & step-by-step resolution