Pular para o conteúdo principal
GE Healthcare · CTLightSpeedBrightSpeed244631244634

LightSpeed 5.X (60kW)

Symptom

Flash download fails repeatedly during a software load, the GTCB or TCB fails to flash.

Error message

Error: 244631 Flash Download(Download) has been failed because of timeout. See Long Description for complete error. Fri Aug 25 11:22:31 2006 Error: 244631 Host: pwct Process: fwmgr File: fwmgr_events.c Line: 9386 Running Version Management Tool Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 Flash Download(Download) has been failed because of timeout. Firmware File: /usr/g/ss_fw/table/tcb.bin Note: Last line above is telling what unit did not download. Firmware File: /usr/g/ss_fw/table/tcb.bin is the TCB board Firmware File: /usr/g/ss_fw/jedi/sc_package.mx is the Smart Cathode Tank on an HD750 Firmware File: /usr/g/ss_fw/jedi/appl.mx

Error: 244634 Subsystem State; DCB:up, CCB:up, -TCB:down -, JEDI:up, TGP/ORP:up TCB:DOWN See Long Description Sun Aug 29 11:13:58 2004 Error: 244634 Host: erct Process: fwmgr File: fwmgr utils.c Line: 660 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 Subsystem State; DCB:up CCB:up TCB:down JEDI:up TGP/ORP:up.

Applies to

Models
LightSpeed, BrightSpeed
Product
LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution