Pular para o conteúdo principal
GE Healthcare · MRSigna2254051

Pre Scan Failed RF Amplifier Not Ready

Symptom

Auto Prescan failed, RF Amplifier is not ready Getting another message that RF Amplifier is powering but message never goes away.

Error message

2254051 Power Monitor 1 Fault. RF Transmit power has reached safety limit allowed for scan parameters and patient weight. See Long Description Fri Aug 20 08:33:11 2004 Error: 2254051 Host: lx-mr Process: NSP File: SCP:PmMdsHWO Line: 1019 Power Monitor 1 Fault. RF Transmit power has reached safety limit allowed for scan parameters and patient weight. Check patient weight entered. amplitude: 35 pulse width: 33 duty cycle: 09 Fault in head amplitude. This could also occur when power cycling SSM.

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Component
2160200-5 CPD Board
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues