GE Healthcare · MRSigna2254051
Pre Scan Failed RF Amplifier Not Ready
Symptom
Auto Prescan failed, RF Amplifier is not ready Getting another message that RF Amplifier is powering but message never goes away.
Error message
2254051 Power Monitor 1 Fault. RF Transmit power has reached safety limit allowed for scan parameters and patient weight. See Long Description Fri Aug 20 08:33:11 2004 Error: 2254051 Host: lx-mr Process: NSP File: SCP:PmMdsHWO Line: 1019 Power Monitor 1 Fault. RF Transmit power has reached safety limit allowed for scan parameters and patient weight. Check patient weight entered. amplitude: 35 pulse width: 33 duty cycle: 09 Fault in head amplitude. This could also occur when power cycling SSM.